| Type of Product: | Device |
| Medsafe Reference: | 28930 |
| Brand Name: | CT Lightspeed Family |
| Model: | CT, Revolution CT ES, Revolution Apex, and Revolution CT with Apex Edition Systems |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | The manufacturer has become aware of an issue where the accumulated dose is incorrectly displayed in certain situations. This occurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/2/2022 |