Detail of Recall

Type of Product:Device
Medsafe Reference:28930
Brand Name:CT Lightspeed Family
Model:CT, Revolution CT ES, Revolution Apex, and Revolution CT with Apex Edition Systems
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Issue:The manufacturer has become aware of an issue where the accumulated dose is incorrectly displayed in certain situations. This occurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/2/2022