Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28911
Brand Name:Blood/Fluid Warming System
Model:3M™ Ranger™ Blood/ Fluid Warming Unit, Model 245
Affected:
Software version:
Recalling Organisation:3M New Zealand Ltd
250 Archers Road
Glenfield
AUCKLAND 0627
Contact Information:0800 808 182
Manufacturer:3M Company 3M Healthcare dba 3M Consumer Health Care, 3M Center
Brand Name:Bair Hugger® Temperature Management System - forced air warming unit
Model:3M™ Bair Hugger™ Warming Unit, Model 775
Affected:
Software version:
Recalling Organisation:3M New Zealand Ltd
250 Archers Road
Glenfield
AUCKLAND 0627
Contact Information:0800 808 182
Manufacturer:3M Company 3M Healthcare dba 3M Consumer Health Care, 3M Center
Issue:The manufacturer has identified that the 3M Ranger Blood/ Fluid Warming Units, Model 245 (Catalogue #24503) and 3M Bair Hugger Warming Units, Model 775 (Catalogue #77503) were supplied and distributed to customers with non-compliant power cords that does not meet AS/NZS 3112 requirements. The supplied power cords are missing the electrical pin insulation on the plug.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/1/2022
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /