Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28911 |
| Brand Name: | Blood/Fluid Warming System |
| Model: | 3M™ Ranger™ Blood/ Fluid Warming Unit, Model 245 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | 3M New Zealand Ltd 250 Archers Road Glenfield AUCKLAND 0627 |
| Contact Information: | 0800 808 182 |
| Manufacturer: | 3M Company 3M Healthcare dba 3M Consumer Health Care, 3M Center |
| Brand Name: | Bair Hugger® Temperature Management System - forced air warming unit |
| Model: | 3M™ Bair Hugger™ Warming Unit, Model 775 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | 3M New Zealand Ltd 250 Archers Road Glenfield AUCKLAND 0627 |
| Contact Information: | 0800 808 182 |
| Manufacturer: | 3M Company 3M Healthcare dba 3M Consumer Health Care, 3M Center |
| Issue: | The manufacturer has identified that the 3M Ranger Blood/ Fluid Warming Units, Model 245 (Catalogue #24503) and 3M Bair Hugger Warming Units, Model 775 (Catalogue #77503) were supplied and distributed to customers with non-compliant power cords that does not meet AS/NZS 3112 requirements. The supplied power cords are missing the electrical pin insulation on the plug. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 31/1/2022 |