Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28872 |
| Brand Name: | ST80i Stress Test System |
| Model: | Philips ST80i Stress Test System - software version A.03.01 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | The manufacturer has identified that when patient demographic information is entered into the system and the patient information is saved or exported to a folder, in rare situations, the demographic information that is displayed is that of the previous patient instead of the current patient resulting in patient data mismatch. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/1/2022 |