Type of Product:Device
Medsafe Reference:28872
Brand Name:ST80i Stress Test System
Model:Philips ST80i Stress Test System - software version A.03.01
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer has identified that when patient demographic information is entered into the system and the patient information is saved or exported to a folder, in rare situations, the demographic information that is displayed is that of the previous patient instead of the current patient resulting in patient data mismatch.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/1/2022