Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28827 |
| Brand Name: | RaySearch_Radiation therapy treatment planning system |
| Model: | RayStation/RayPlan 6, 7, 8A, 8B, 9A, 9B including some service packs |
| Affected: | |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | The manufacturer has identified an incorrect display of patient images in RayStation during treatment planning. It concerns an issue found with the display of secondary image sets in a side-by-side view in RayStation/RayPlan 6, 7, 8A, 8B, 9A, 9B including some service packs. The primary image set may be displayed instead of the secondary image set. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/1/2022 |