Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28780
Brand Name:Veradius/ Veradius Unity/ BV Systems/ Zenition
Model:
Affected:Wireless Foot switch 3P - Model number 459801238191 and 459801238231
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has discovered that the wireless foot switch can suddenly stop responding due to a firmware issue when a number of ambient conditions coexist, such as EMC disturbance and the presence of other wireless devices in the room.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/1/2022