Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28760 |
| Brand Name: | ImmunoCAP Total IgE, ImmunoCAP Specific IgE and ViewAllergy |
| Model: | |
| Affected: | Multiple Lot numbers |
| Software version: | |
| Recalling Organisation: | Abacus dx New Zealand Unit B, 30 Vestey Drive Mt Wellington Auckland 1060 |
| Contact Information: | 0800222170 |
| Manufacturer: | Thermo Fisher Scientific Microbiology Basingstoke |
| Issue: | The manufacturer has seen an increased frequency of dispensing errors on the Phadia platforms due to capsules stuck in carriers during Q3 2021. Small deformations of capsules may occasionally occur during manufacturing. It has been identified that one piece of the production equipment has been damaging the capsules intermittently over a period of time. The deformation is located underneath the flange of the capsule and may cause the capsules to get stuck inside the carriers. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/12/2021 |