Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28743
Brand Name:AU/DxC AU Magnesium
Model:OSR6189
Affected:DxC AU Magnesium
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Ireland Inc
Issue:The manufacturer has determined that the Magnesium failed to meet the lipemic interference specification for the serum application as defined in the IFU: Interference less than 10% up to 500 mg/dL Intralipid.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/12/2021