Detail of Recall

Type of Product:Device
Medsafe Reference:28736
Brand Name:AU/DxC AU Uric Acid - Uric acid IVD, kit, enzyme spectrophotometry
Model:OSR6x98
Affected:2571, 2566, 2623, 2552, 2555, 2553, 2548, 2551, 2548, 2554, 2574
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Ireland Inc
Issue:The manufacturer has identified that Uric Acid did not meet the interference acceptance criteria for the low and high pool for the conjugated bilirubin interference claim. It generates Incorrect Results.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/12/2021