Detail of Recall

Type of Product:Device
Medsafe Reference:28734
Brand Name:InteleradPACS (Intelerad Medical Systems Incorporated)
Model:IntelePACS
Affected:
Software version:Enhanced Viewer software (part of IntelePACS software), from version 1.17.4 to version 1.19.19. -InteleViewer software (part of IntelePACS software), from version 4-3-1 R1 to version 5-4-1
Recalling Organisation:Pharmaceutical Marketing New Zealand Ltd
162 Nile Road
Milford
Auckland 0620
Contact Information:09 377 0660
Manufacturer:Intelerad Medical Systems Incorporated
Brand Name:Intelerad-35719-Class IIa-InteleradPACS
Model:
Affected:
Software version:Enhanced Viewer software (part of IntelePACS software), from version 1.17.4 to version 1.19.19. -InteleViewer software (part of IntelePACS software), from version 4-3-1 R1 to version 5-4-1
Recalling Organisation:Clinical & Regulatory Services Limited
30 Fulford Road
Havelock North
Hastings 4130
Contact Information:(06) 844 4490
Manufacturer:Intelerad Medical Systems Incorporated
Issue:The manufacturer advised that when using advanced visualisation to reconstruct MR, CT and PET images that contain non-square pixels, the reconstructed image is incorrect. Non-square pixels occur when the row and column pixel spacing values are different.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/12/2021