Detail of Recall

Type of Product:Device
Medsafe Reference:28664
Brand Name:TaqPath
Model:
Affected:Catalog #A47814
Software version:
Recalling Organisation:Thermo Fisher Scientific New Zealand Ltd
244 Bush Road
Albany
AUCKLAND
Contact Information:0800 636 327
Manufacturer:Life Technologies AS
Issue:The manufacturer has identified that when the TaqPathTM COVID-19 Combo Kit, 1000 reactions is used with the QuantStudio 5 there is a remote occurrences (~0.0002%) of signal discontinuity in the multicomponent plots from QuantStudio 5 (QS5 and QS5 Dx) instruments, for 96-well, 0.2-mL block and a MicroAmp™ Optical 96-Well Reaction Plate (with or without barcode). The investigation has determined that the observed signal discontinuities are likely caused by inadequate sealing of the PCR plate.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/12/2021