Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28664 |
| Brand Name: | TaqPath |
| Model: | |
| Affected: | Catalog #A47814 |
| Software version: | |
| Recalling Organisation: | Thermo Fisher Scientific New Zealand Ltd 244 Bush Road Albany AUCKLAND |
| Contact Information: | 0800 636 327 |
| Manufacturer: | Life Technologies AS |
| Issue: | The manufacturer has identified that when the TaqPathTM COVID-19 Combo Kit, 1000 reactions is used with the QuantStudio 5 there is a remote occurrences (~0.0002%) of signal discontinuity in the multicomponent plots from QuantStudio 5 (QS5 and QS5 Dx) instruments, for 96-well, 0.2-mL block and a MicroAmp™ Optical 96-Well Reaction Plate (with or without barcode). The investigation has determined that the observed signal discontinuities are likely caused by inadequate sealing of the PCR plate. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 1/12/2021 |