Type of Product:Device
Medsafe Reference:28646
Brand Name:DS Maref SIPC
Model:
Affected:Serial numbers: 220112944 to 20117954 and 2026322 to 2032792
Software version:
Recalling Organisation:Medirent NZ Ltd
C/O Deloitte
80 Queens Street
Auckland 1010
Contact Information:
Manufacturer:Daesung Maref Co Ltd
Issue:Through improved testing it was observed in a number of device that as the devices cycles through the treatment the pressure delivered to the garment gradually increases beyond what the device is set to. This pressure can become excessive for the user and exceeds the devices pressure specifications.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/11/2021