Detail of Recall

Type of Product:Device
Medsafe Reference:28636
Brand Name:BabyLeo TN500 Incuwearmer
Model:
Affected:Drager Babyleo TN500 IncuWarmer installed with software lower than 1.06.00
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Issue:The manufacturer has identied that devices equipped with the Auto-Thermo Package, the Weaning procedure helps to automatically reduce the air temperature in a certain time interval defined according to the hospital protocol to wean patients from the incubator to an open care device.
In recent software versions, the weaning function was stopped when Kangaroo Mode was activated. Consequently, the workflow was interrupted, and the weaning procedure on Babyleo TN500 had to be started again.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory