Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28624
Brand Name:Veradius/ Veradius Unity/ BV Systems/ Zenition
Model:Multiple model numbers
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:Angiographic X-Ray
Model:Multiple model numbers
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:Azurion MDR
Model:Multiple model numbers
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has discovered that the wireless foot switch can suddenly stop responding due to a firmware bug when a number of ambient conditions coexist, such as EMC disturbance and the presence of other wireless devices in the room. If the wireless foot switch fails, fluoroscopy imaging and exposure will not be available until a cold restart of the system [power down, power up] is performed.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/11/2021