Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28624 |
| Brand Name: | Veradius/ Veradius Unity/ BV Systems/ Zenition |
| Model: | Multiple model numbers |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Brand Name: | Angiographic X-Ray |
| Model: | Multiple model numbers |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Brand Name: | Azurion MDR |
| Model: | Multiple model numbers |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Issue: | The manufacturer has discovered that the wireless foot switch can suddenly stop responding due to a firmware bug when a number of ambient conditions coexist, such as EMC disturbance and the presence of other wireless devices in the room. If the wireless foot switch fails, fluoroscopy imaging and exposure will not be available until a cold restart of the system [power down, power up] is performed. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/11/2021 |