Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28601 |
| Brand Name: | InteleradPACS |
| Model: | |
| Affected: | IntelePACS software - Enhanced Viewer software (part of IntelePACS software), from version 4.16.1.EV.1.5.5 to 4.16.1.EV.1.19.14 |
| Software version: | |
| Recalling Organisation: | Pharmaceutical Marketing New Zealand Ltd 162 Nile Road Milford Auckland 0620 |
| Contact Information: | 09 377 0660 |
| Manufacturer: | Intelerad Medical Systems Incorporated |
| Issue: | The manufacturer has validated that your IntelePACS software could potentially be affected. The issue may be triggered by the presence of pre-formatted HTML text in HL7 reports using the tag. These reports will be truncated when they are rendered by the Enhanced Viewer PDF report viewer, which could lead to incomplete report content displayed to the Enhanced Viewer user. This issue will only happen while using Microsoft Edge/Chromium 92, Mozilla Firefox 90, Google Chrome 92 and later versions of these browsers. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/11/2021 |