Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28598 |
| Brand Name: | Vscan |
| Model: | Vscan Extend - Battery Depletion |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Vingmed Ultrasound AS |
| Issue: | The manufacturer has become aware of an issue where the Vscan Extend will not power on because of a depleted internal battery. This issue can occur if the device is not turned off prior to the external battery being removed or when connected to a Personal Computer (PC) without connecting to AC power. In these situations, the device is not available for use until it is connected to AC power and charged for approximately 10 minutes. This could cause a delay in treatment. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/11/2021 |