Detail of Recall

Type of Product:Device
Medsafe Reference:28586
Brand Name:QIAstat-Dx Gastrointestinal Panel
Model:
Affected:cat. no. 691411
Software version:
Recalling Organisation:Qiagen GmbH
Qiagen Strasse 1
Hilden 40724
Germany
Contact Information:
Manufacturer:Qiagen GmbH
Issue:The manufacturer has become aware that the reaction chambers contain incorrect primer–probe mixes. If an affected cartridge is used, several possibilities for false-negative, false-positive, or a combination of both false-negative and false-positive results could occur.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/11/2021