Type of Product:Device
Medsafe Reference:28566
Brand Name:Anaesthesia Work Stations
Model:Carestation 750/750c anesthesia delivery systems
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has become aware of an issue where the Carestation 750/750c Anesthesia Delivery Systems - after a specific sequence of actions, audible tone of O2 supply pressure low alarm will remain silenced until system reboot.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/11/2021