Detail of Recall

Type of Product:Device
Medsafe Reference:28501
Brand Name:Freelite Kappa/Lambda
Model:
Affected:Optilite Freelite® Kappa Free Kit (LK016.OPT) - lot 470792, Optilite Freelite Kappa Free Kit (x10) (LK016.10.OPT) - lot 470794, Optilite Freelite Mx™ Kappa Free Kit (LK016.M.OPT) - lot 473522
Software version:
Recalling Organisation:The Binding Site Limited
Grant Thornton New Zealand Ltd
L4, 154 Fanshawe Street
Auckland 1010
Contact Information:
Manufacturer:The Binding Site Group Ltd
Issue:The manufacturer has identified that the kits are likely to exhibit AC flags on calibrator 7. An AC flag occurs when the reaction kinetics exceed the Optilite threshold limits for the prozone check. When this occurs on a calibrator the flag is shown in the 'errors' field within the F4>3 calibration results screen of the Optilite software.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/10/2021