Type of Product:Device
Medsafe Reference:28500
Brand Name:Freelite Kappa/Lambda
Model:
Affected:LK018.S - lot 472776, LK018.10S - lot 472664, LK018.L.S - lot 435251
Software version:
Recalling Organisation:The Binding Site Limited
Grant Thornton New Zealand Ltd
L4, 154 Fanshawe Street
Auckland 1010
Contact Information:
Manufacturer:The Binding Site Group Ltd
Issue:It has been identified that calibration curves produced using LK018.S lot 472776 have a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 (110.45mg/L) and calibrator point 6 (173.76mg/L), which can be exacerbated when calibrator 6 activity reaches the optical ceiling of the SPA+ analyser.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/10/2021