Detail of Recall

Type of Product:Device
Medsafe Reference:28467
Brand Name:ARCHITECT Systems
Model:ARCHITECT i1000SR, ARCHITECT i2000SR, ARCHITECT i2000, ARCHITECT c4000, ARCHITECT c8000, ARCHITECT c16000
Affected:
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd
Unit D
4 Pacific Rise
Mt Wellington
Auckland 2025
Contact Information:0800 656 233
Manufacturer:Abbott Laboratories
Issue:The manufacturer has identified three potential performance issues for the ARCHITECT Software version 9.41 and earlier. Abbott is releasing ARCHITECT Software versions 9.45 and 9.50 to correct these issues.
During the investigation, more issues have been found, such as: 1. Unable to process test, internal wash pressure (x) error (y) pipettor'. 2. Carryover of incorrect calibrator sample volumes. 3. i-Automated Reconstitution Module (iARM) loses communication with the System Control Center (SCC). 4. Erroneous ICT calibration curves. 5. Quality control (QC) data are not deleted, and calibration curves remain active. 6. 1ml Syringes wear over time. 7. Control flags are missing. 8. Signal spikes during the read cycle. 9. Abnormal optics reads. 10. Tracking multiple RVC errors. 11. Incorrect aspiration issue. 12. Inconsistent rounding of configured linearity values.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/10/2021