Detail of Recall

Type of Product:Device
Medsafe Reference:28447
Brand Name:Glucose reagent OSR6x21
Model:
Affected:All lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has determined that the Glucose Haemolysate application did not consistently meet the current haemolytic interference claim i.e. interference less than 20% up to 150 g/L haemoglobin. The low analyte concentration pool exhibited a maximum negative bias of -21.24% at 150 g/L haemoglobin.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/10/2021