Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28425
Brand Name:Thermometer, infrared, ear
Model:Braun ThermoScan® PRO 6000 Ear Thermometer.
Affected:
Software version:
Recalling Organisation:Ebos Group Ltd
108 Wrights Road
Addington
Christchurch
Contact Information:0800 18 17 16
Manufacturer:Welch Allyn Inc
Issue:The manufactuere has identified a potential risk associated with overheating of the Braun ThermoScan PRO 6000 probe tip due to fluid ingress. If the device is exposed to fluid ingress and is used before the cleaning fluid has completely dried, there is a risk of the device overheating potentially causing a burn to the user or the ear canal of the patient.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/10/2021