| Type of Product: | Device |
| Medsafe Reference: | 28423 |
| Brand Name: | PrisMax |
| Model: | |
| Affected: | Product Code - 955558 All serial numbers with software versions 2.x |
| Software version: | |
| Recalling Organisation: | Baxter Healthcare Ltd 33 Vestey Drive Mt Wellington AUCKLAND |
| Contact Information: | 0800 229 837 |
| Manufacturer: | Baxter Healthcare SA (Switzerland) |
| Issue: | The manufacturer has identified a software anomaly occurring during use. If the operator initiates therapy with a saved prescription profile the PrisMax System may display values from the original prescription profile, rather than the current prescription. On the PrisMax System interface, the user is always asked to confirm all prescription values before initiating therapy or when modifying the prescription during therapy. 05/11/2021: Interim Report |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/10/2021 |