Detail of Recall

Type of Product:Device
Medsafe Reference:28420
Brand Name:Oncentra RTP
Model:
Affected:Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Nucletron BV
Issue:The manufacturer has identified in two rare scenarios, that the applicator model may get mirrored. This is caused by a design fault in the mapping of the applicator model to the user provided anchor points. If unnoticed, this could result in a difference between the planned and delivered dose.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/10/2021