Detail of Recall

Type of Product:Device
Medsafe Reference:28416
Brand Name:Balloon pump
Model:
Affected:All Serial numbers
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Datascope Corporation
Issue:This notification is an update to the previously initiated action, dated 4 October 2021 regarding Helium overreporting and Battery removal shutdown.
The manufacturer has determined that in the presence of some radiated RF disturbances, the Helium Indicator on the Cardiosave IABP would intermittently over-report the remaining Helium Capacity. And Shutdown Upon Battery Removal.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/3/2024