Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28390
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model:AppSW VD20C/VD20N
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:The manufacturer has identified that the software can produce an incorrect interpretation of the table rotation if the table has been rotated by more than 4° in either direction. This can cause the 3D image reconstruction to be shown rotated by up to 5° relative to the patient and to be used inappropriately by the system.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/9/2021