Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28389 |
| Brand Name: | TiLite Wheelcahirs Fixed Frame |
| Model: | |
| Affected: | manufactured between January 19, 2017 and June 14, 2021 with the serial number range from M2005054 to 1231112011 |
| Software version: | |
| Recalling Organisation: | Permobil NZ Ltd 181A McLeod Road Waitakere City Auckland 0650 |
| Contact Information: | 0800 115 222 |
| Manufacturer: | Permobil AB |
| Issue: | The manufacturer has identified that the transit tie downs on the TiLite AERO Z wheelchair model must be configured in the inward and rearmost position in order to pass ANSI/RESNA WC19 or ISO 7176-19 testing. Therefore, the transit tie downs on the TiLite AERO Z must be configured in this specific orientation for the wheelchair to be used as a passenger seat in a motor vehicle. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/9/2021 |