Detail of Recall

Type of Product:Device
Medsafe Reference:28389
Brand Name:TiLite Wheelcahirs Fixed Frame
Model:
Affected:manufactured between January 19, 2017 and June 14, 2021 with the serial number range from M2005054 to 1231112011
Software version:
Recalling Organisation:Permobil NZ Ltd
181A McLeod Road
Waitakere City
Auckland 0650
Contact Information:0800 115 222
Manufacturer:Permobil AB
Issue:The manufacturer has identified that the transit tie downs on the TiLite AERO Z wheelchair model must be configured in the inward and rearmost position in order to pass ANSI/RESNA WC19 or ISO 7176-19 testing. Therefore, the transit tie downs on the TiLite AERO Z must be configured in this specific orientation for the wheelchair to be used as a passenger seat in a motor vehicle.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/9/2021