Database updated: 3 November 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28349
Brand Name:Conductor Gel
Model:
Affected:Gel, ultrasonic coupling
Software version:
Recalling Organisation:DJO Global
Contact Information:
Manufacturer:Eco-Med Pharmaceutical Inc.
Brand Name:Shockwave
Model:
Affected:DJO kitted products with gel or lotion as a sample
Software version:
Recalling Organisation:DJO Global
Contact Information:
Manufacturer:DJO LLC
Brand Name:Therapeutic Ultrasound
Model:
Affected:DJO kitted products with gel or lotion as a sample
Software version:
Recalling Organisation:DJO Global
Contact Information:
Manufacturer:DJO LLC
Brand Name:Chattanooga RPW 2 (2174)
Model:
Affected:DJO kitted products with gel or lotion as a sample
Software version:
Recalling Organisation:DJO Global
Contact Information:
Manufacturer:DJO France SAS
Brand Name:Chattanooga Intelect Mobile 2
Model:
Affected:DJO kitted products with gel or lotion as a sample
Software version:
Recalling Organisation:DJO Global
Contact Information:
Manufacturer:DJO France SAS
Brand Name:Chattanooga Intelect Mobile 2 Combo
Model:
Affected:DJO kitted products with gel or lotion as a sample
Software version:
Recalling Organisation:DJO Global
Contact Information:
Manufacturer:DJO France SAS
Issue:The manufacturer is advising that all users should immediately stop use and dispose of ALL CHATTANOOGA ultrasound gels and Myossage lotions, regardless of lot, due to the risk of bacterial contamination with Burkholderia cepacia complex (Bcc). Affected gels or lotions have been supplied individually or as part of a DJO kitted product.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/9/2021
Hide menus