Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28346
Brand Name:EMLA
Model:
Affected:Digital Linear Accelerator
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Solutions AB
Issue:The manufacturer has identified a risk that the Diode D1 in the Modulator can fail in a destructive manner once removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter. As the destructive event results in the release of silicon oil and ceramic fragments, further mitigating actions need to be taken when dealing with the diode failure, to reduce the risk of injury to users in the immediate vicinity.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/9/2021