Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28319 |
| Brand Name: | W/A Thermometer, infrared, ear |
| Model: | all models |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | Welch Allyn Inc |
| Issue: | The manufacturer has identified a potential risk associated with overheating of the Braun ThermoScan PRO 6000 probe tip due to fluid ingress. This issue potentially impacts all PRO 6000 in the field. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/9/2021 |