Detail of Recall

Type of Product:Device
Medsafe Reference:28274
Brand Name:Inorganic Phosphorus
Model:
Affected:All lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:During internal interference testing, Inorganic Phosphorus failed to meet the hemolytic interference specification for the serum application. The low analyte concentration pool tested (~0.7 mmol/L) exhibited a positive bias with a maximum bias of 24.02% at 3.5 g/L haemolysate. This positive bias is contributed to by phosphorus release from haemolysed cells. Current product labeling reflects only optical interference from haemoglobin and does not account for Inorganic Phosphorus released from haemolysed cells. There is no hemolysis issue with the urine application.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/9/2021