Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28267 |
| Brand Name: | ATLAN Family of Anaesthesia Machines |
| Model: | |
| Affected: | with software 1.01 and lower |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Issue: | The manufacturer has become aware of cases in which the Atlan Anesthesia Workstation with sw 1.01 has shown artefacts on the pressure and flow curves while the devices are being operated in one of the following ventilation modes: PC - CMV, PC - SIMV, PC - SIMV / PS, VC - CMV / Autoflow, VC - SIMV / Autoflow, VC - SIMV / PS / Autoflow, CPAP / PSV. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/8/2021 |