Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28248 |
| Brand Name: | Infinity Acute Care System Workstation Critical Care |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Brand Name: | Babylog VN500 Ventilator |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Issue: | A restart is a designed system recovery function to restore the proper function of the ventilator. In the rare reported cases, however, an error in the data processing of the activated CO2 measurement caused the restart of the ventilation unit. Devices without integrated or with disabled CO2 measurement function are not affected. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/8/2021 |