Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28237
Brand Name:SureSigns
Model:Philips SureSigns VS4 Vital Signs - All serial numbers
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer has received reports of VS4s failing to exit the Standby mode. This means that the monitor doesn't restart when the user presses the On/Standby key. This happens when the monitor has been left in Standby mode for an extended period. For example, when left in Standby mode overnight while the battery is charging.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/8/2021