Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28201
Brand Name:Anaesthesia absorber cartridge
Model:
Affected:Multiple Product codes
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Armstrong Medical Ltd
Issue:The manufacturer is aware of reports indicating that a small number of devices are associated with high and unexpected resistance to gas flow during clinical use. Such resistance could impact or prevent adequate ventilation of an anaesthetised patient, should a defective device go into clinical use.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/8/2021