Detail of Recall

Type of Product:Device
Medsafe Reference:28193
Brand Name:CombiDiagnost R90
Model:Model number - 709030 - software - R1.0, R1.1
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The manufacturer has become aware of a software issue, where the kV and mA values are not locked but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image. However, if the user stops and restarts fluoroscopy, then the Lock-In works as specified.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/8/2021