Detail of Recall

Type of Product:Device
Medsafe Reference:28175
Brand Name:Access hsTnI Reagent
Model:
Affected:All Lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:A subsequent investigation has determined that sample-to-sample carryover may also occur under certain conditions confirming that intra-assay carryover encompasses into-pack and sample-to-sample carryover. Through these subsequent studies, BEC determined that clinically significant sample-to-sample carryover can occur in hsTnI samples that are tested after a sample with a cTnI concentration >55,000 pg/mL (ng/L).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/2/2022