Type of Product:Device
Medsafe Reference:28169
Brand Name:Veradius/ Veradius Unity/ BV Systems/ Zenition
Model:Allura Xper systems with software releases R7.6.x, R7.7.x, R7.8.x, =R8.1.17.x, =R8.2.17.x
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has discovered through customer feedback and internal testing an intermittent issue with the device software that affects the timing of displayed information and the system's response to user commands. This issue is due to a blockage of the Microsoft Windows EventLog on the host computer for a period of up to approximately 25 seconds, with most delays between 8 and 15 seconds.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/8/2021