Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:28144
Brand Name:VIROTROL ToRCH-M Class B
Model:
Affected:Lot number - 128720
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:Virotrol Torch batch 128720 was slightly non-reactive for Toxo IgG at finished product testing. The analyte was declaimed for this batch.
The package insert for the Lot # 128720 will indicate that no claims are made for stability or performance of TOXO IgG. A notification sticker is attached on the product box label for customers.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/8/2021