Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28144 |
| Brand Name: | VIROTROL ToRCH-M Class B |
| Model: | |
| Affected: | Lot number - 128720 |
| Software version: | |
| Recalling Organisation: | Bio-Rad New Zealand Level 30 48 Shortland Street AUCKLAND 1010 |
| Contact Information: | 0508 805 500 |
| Manufacturer: | Bio-Rad Laboratories Inc |
| Issue: | Virotrol Torch batch 128720 was slightly non-reactive for Toxo IgG at finished product testing. The analyte was declaimed for this batch. The package insert for the Lot # 128720 will indicate that no claims are made for stability or performance of TOXO IgG. A notification sticker is attached on the product box label for customers. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/8/2021 |