Type of Product:Device
Medsafe Reference:28122
Brand Name:Olympus - Caps (Japan)
Model:model MAJ-891 (MAJ-891)
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The manufacturer is investigating potential causes and contributing factors relating to reported infection and contamination complaints associated with the MAJ-891. Although the manufacturer has determined that the risk of infection is low based on available data, investigation revealed a potential risk of patient infection due to inadvertent retention of biomaterial when the reprocessing procedure applicable to the MAJ-891 is not performed as instructed, such as not completely removing and disassembling MAJ-891 during reprocessing
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/7/2021