Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 28060 |
| Brand Name: | LIKO Likoguard Ceiling Hoist System |
| Model: | LikoGuard L (3301030), LikoGuard XL (3301040) and Emergency stop/lowering strap 33090020 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | Liko AB |
| Issue: | The manufacturer has become aware of a potential issue related to the length of the Emergency strap on overhead lifts LikoGuard L and XL. Through the investigation, it was identified that the Emergency strap could be longer than the intended length after installation of the overhead lift. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/7/2021 |