Detail of Recall

Type of Product:Device
Medsafe Reference:28013
Brand Name: COLONOSCOPES
Model:
Affected:All serial number
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:CV-170 Video Image Processor
Model:
Affected:All serial numbers
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The manufacturer has now completed developing countermeasures for the CV-1500 processor to mitigate unintentional thermal tissue damage when used in conjunction with EZ1500 series endoscopes. With a new software update, the maximum light intensity will be improved and better controlled, without negative impact on the image quality.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/7/2021