Detail of Recall

Type of Product:Device
Medsafe Reference:27974
Brand Name:V60 / V60 PLUS Ventilator
Model:
Affected:Software Version 3.00 or Software Version 3.10
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics California LLC
Issue:The manufacturer has identified that when V60 ventilators with High Flow Therapy option upgrade and V60 Plus ventilators reach the maximum high flow therapy pressure limit, the Cannot Reach Target Flow alarm will sound and the flow rate will decrease in order to decrease the pressure. In some cases, patients may experience oxygen desaturation as a result of the decrease in flow rate, which can be characterized as moderate or severe hypoxemia
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/7/2021