Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27973
Brand Name:Soltive System
Model:TFL-PLS SOLTIVE Premium SuperPulsed Laser System - All Serial numbers
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Gyrus ACMI Inc
Issue:The manufacturer has received reports of patient injury associated with the Soltive Laser System, when used in dusting and fragmentation of ureteral stones. The investigations revealed a user likely exceeded the 20W standard presets for stone dusting and fragmentation when treating impacted stones in the ureter and ureteropelvic junction.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/7/2021
Hide menus
Show menus