Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27959 |
| Brand Name: | ThermoScientificTM OxoidTM AMC30 Amoxycillin / Clavulanic Acid Antimicrobial |
| Model: | |
| Affected: | 2394162 29/08/2021, 2511707 24/04/2022, 2609976 16/07/2022, 2832507 05/08/2022 |
| Software version: | |
| Recalling Organisation: | Thermo Fisher Scientific New Zealand Ltd 244 Bush Road Albany AUCKLAND |
| Contact Information: | 0800 636 327 |
| Manufacturer: | |
| Issue: | An internal investigation has confirmed that lots of CT0223B, ThermoScientificTM OxoidTM Amoxycillin/Clavulanic Acid Antimicrobial Susceptibility Discs, are giving small zones of inhibition for Quality Control organism Escherichia coli ATCC®35218. The zones of inhibition are outside the specified CLSI/EUCAST 17-22 mm limits. The issue is suspected to be caused by differing levels of residual moisture in the product leading to faster rates of degradation of the primary antibiotic and the beta-lactamase inhibitor. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/6/2021 |