Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27919
Brand Name:STA R Max® analyzer
Model:
Affected:Software version 4.06 on STA R Max
Software version:
Recalling Organisation:Diagnostica Stago Pty Ltd
Level 3
64 Cashel Street
Christchurch
Contact Information:+64 3 377 4115
Manufacturer:Diagnostica Stago SAS
Issue:Internal investigation has identified an issue on special or special plus washes with STA-Desorb U on reagent needles (#2 and #3) when a level detection error (LLD) occurs. This washing anomaly is most commonly present in a specific context of non-recommended use (unloading and reloading of bottles of reagents already used, with entry of an incorrect residual volume by the user).
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/6/2021