Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27918
Brand Name:Endocuff
Model:
Affected:Lots - 60496 to 70914
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Boddingtons Plastics Ltd
Issue:Due to an irregularity in the packaging process, it has been identified that there is a very low probability for a potential packaging seal defect which may allow a breach of the package's sterile barrier. This seal breach can be detected with the naked eye prior to use once the lid has been removed from the plastic package. A breach of the package's sterile barrier might lead to patient infection.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/6/2021
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /