Detail of Recall

Type of Product:Device
Medsafe Reference:27914
Brand Name:ORTHO VISION
Model:ORTHO VISION Analyzer for BioVue and ORTHO VISION Max Analyzer for BioVue
Affected:
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics
c/o Quigg Partners
Level 7, 36 Brandon Street
Wellington 6011
Contact Information:0800 402 899
Manufacturer:Ortho-Clinical Diagnostics (UK)
Issue:The maufacturer has identified that the ORTHO VISION™/ORTHO VISION™ Max analyzer, while in maintenance mode, a specific operator action (pressing the cancel button while in a maintenance task) may cause unexpected movement of the grip arm and pipettor. Pressing the cancel button during maintenance is not typically done as this action results in a failed maintenance task and the need to repeat the task in its entirety before resuming processing.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/6/2021