Detail of Recall

Type of Product:Device
Medsafe Reference:27907
Brand Name:Floseal Applicators_47187_BS Zurich
Model:
Affected:0000357425
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Baxter Healthcare Corporation SA
Issue:The manufacturer has identified that the expiry date which is encoded into the 2D barcode located on the primary and secondary package labeling does not match the correct printed expiry date on the primary and secondary package labeling. The human-readable expiry dates printed on the primary and secondary package labeling are in the YYYY-MM-DD format, whereas the 2D barcode expiry date was encoded as the YY-DD-MM format.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/6/2021