Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27891
Brand Name:CS - CoreValve Evolut R_Low Risk
Model:EnVeo PRO Delivery Catheter System - ENVPRO-14, ENVPRO-16, ENVEOR-N
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Core Valve LLC
Brand Name:CS - CoreValve Evolut PRO_Low Risk
Model:
Affected:Multiple lots
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Core Valve LLC
Brand Name:SH - Medtronic Evolut™ PRO+ System
Model:
Affected:Multiple lots
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Core Valve LLC
Issue:The manufacturer is recalling EnVeo™ R, EnVeo™ PRO and Evolut™ PRO+ Delivery Catheters with a manufacturing date before 11-JUL-2020. Due to actuator separation that may occur during valve loading on the delivery system, advancement, deployment or recapture of the valve.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/6/2021
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /